As a professional X-ray machine manufacturer with over 20+ years of factory experience, Fatali-Med follows a strict, traceable quality control system to ensure every X-ray unit delivers stable imaging performance, clinical safety, and long-term reliability.
From component sourcing to final shipment, each system is tested in accordance with IEC and CE medical standards, with full test records available for customer reference. This is how we make sure what you receive performs exactly as promised — not just on day one, but years later.
>> To view more visit: From Fatali's Office to the Workshop in One Shot – Real Footage, Real Strength
For hospitals, clinics, and distributors, X-ray equipment quality is not just about image clarity — it directly impacts:
Diagnostic accuracy
Radiation safety compliance
Equipment lifespan and maintenance cost
Downtime risk and clinical workflow stability
That’s why Fatali-Med treats quality control as a manufacturing system, not a final inspection step.
Every system starts with strict component verification. Key parameters are tested using calibrated medical instruments:
X-ray tube output stability
Generator voltage (kVp) accuracy
Detector sensitivity and response uniformity
All measurements are logged to ensure consistency before assembly begins.
Engineers fine-tune each system under simulated clinical conditions, adjusting:
kVp, mA, and exposure time
Grayscale linearity
Spatial resolution and contrast
This step ensures reliable imaging across different patient sizes and examination types.
Every X-ray unit undergoes a 72-hour non-stop aging test to verify long-term stability.
During this process, we continuously record:
Temperature variation
Voltage stability
Image consistency
Operational noise level
This test helps identify potential failures before the system reaches your site.
Mechanical stability directly affects imaging accuracy. Each system passes:
Stand and table balance tests
Movement smoothness checks
Detector alignment verification
No vibration, drift, or misalignment is accepted.
Before shipment, all systems are tested according to IEC / CE medical safety standards, including:
Leakage current
Grounding resistance
Radiation dose linearity
Only systems that fully pass safety compliance are released.
| Test Item | Standard Tool | Recommended Criteria | Purpose |
|---|---|---|---|
| Tube Output Stability | Dosimeter | ±5% fluctuation | Consistent radiation |
| Voltage Accuracy (kVp) | kVp Meter | ±2 kVp deviation | Generator calibration |
| Exposure Linearity (mAs) | Dosimeter | R² ≥ 0.99 | Exposure accuracy |
| Detector Uniformity | Phantom Image | ≤ ±10% grayscale | Image consistency |
| Aging Test | Data Logger | ≤15°C rise, ≤60 dB | System stability |
| Mechanical Alignment | Level Meter | ≤2 mm deviation | Precision movement |
| Electrical Safety | IEC Tester | ≤0.5 mA leakage | CE/IEC compliance |
| Dose Verification | Dosimeter | ≤ ±10% deviation | Radiation accuracy |
| Image Resolution | Phantom | ≥3.2 lp/mm | Imaging clarity |
| Repeatability | Dosimeter | STD/Mean ≤ 0.05 | Exposure precision |
To help buyers, engineers, and distributors evaluate X-ray systems more confidently, Fatali-Med provides a practical QA inspection checklist based on our factory standards.
This checklist helps you:
Verify system performance during installation
Perform acceptance testing before clinical use
Compare different suppliers objectively
Reduce post-installation risk
👉 Download the Free QA Checklist (PDF)
No spam. Just a practical tool used by our engineers.
During installation or commissioning, customers can repeat key checks — such as image uniformity, exposure repeatability, and mechanical smoothness — using the same QA logic applied in our factory.
Our engineers also provide online technical support, ensuring your team can verify system performance with confidence, even after installation.
20+ years of dedicated X-ray manufacturing
Full IEC / CE-aligned QA documentation
Stable supply chain and component traceability
Proven installations in hospitals and clinics worldwide
Quality is not a marketing promise — it’s a measurable process.
If you’re evaluating suppliers or planning an equipment upgrade, tell us about your application and market requirements.
We’ll recommend the most suitable configuration — backed by 23 years of factory experience and real QA data.
👉 Contact Fatali-Med for a tailored solution & QA report